EMC-Verification
104
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
28
EMC-Verification
The device must not be used directly next to other devices. If this is
required, the device must be continually monitored to ensure its
proper operation under these applied conditions.
Table 1 according to IEC 60601-1-2:2007
Guidelines and manufacturer's declaration - electromagnetic interference
The TruSystem 7000 operating table is intended for use in the environments as specified below. The customer or user of the
aforementioned device should ensure that it is operated in one of the environments as described.
Emitted interference measurements
Compliance
Electromagnetic environment - guidelines
RF emissions in accordance with CISPR 11
Group 1
The TruSystem 7000 operating table exclusively
uses HF energy for its internal FUNCTION.
Therefore, its HF emissions levels are very low. It is
improbable that neighboring devices would be
adversely affected.
RF emissions in accordance with CISPR 11
Class A
The TruSystem 7000 operating table is suitable for
use in establishments other than domestic and those
connected directly to the PUBLIC LOW-VOLTAGE
NETWORK that supplies buildings used for
domestic purposes.
Harmonic Emissions as per IEC 61000-3-2 Class A
Voltage fluctuation/flicker emissions in
accordance with IEC 61000-3-3
Compliant
Table 2 according to IEC 60601-1-2:2007
Guidelines and manufacturer's declaration - electromagnetic immunity
The TruSystem 7000 operating table is intended for use in the environments as specified below. The customer or user of the
aforementioned device should ensure that it is operated in one of the environments as described.
Interference immunity
test
IEC 60601 immunity
Compliance level
Electromagnetic environment -
guidelines
Static electricity
discharge (ESD)
according to
IEC 61000-4-2
±6 kV contact discharge
±8 kV air discharge
±6 kV contact discharge
±8 kV air discharge
Floors should be made from wood or
concrete or covered with ceramic tiles.
If a floor is covered with synthetic
material, the relative humidity must be
at least 30 %.
Rapid transient electrical
disturbances/bursts
pursuant to
IEC 61000-4-4
± 2 kV for mains cables
± 1 kV for input and output
cables
± 2 kV for mains cables
± 1 kV for input and output
cables
The power supply quality should
correspond to that of a typical
commercial or hospital environment.
Voltages (surges) in
accordance with
IEC 61000-4-5
±1 kV outer-conductor/
outer-conductor voltage
±2 kV outer-conductor/
ground voltage
±1 kV outer-conductor/
outer-conductor voltage
±2 kV outer-conductor/
ground voltage
The power supply quality should
correspond to that of a typical
commercial or hospital environment.