e-STUDIO200L/202L/230/232/280/282 TROUBLESHOOTING
June 2004 © TOSHIBA TEC
5 - 54
5.1.8
Scanning system related service call
[C260]
Peak detection error
Does the exposure lamp light? (Perform the output check in the test mode: 03-267)
l
l
l
l
l
l
l
l
l
#
YES
1) Check if the connectors on the CCD and SLG boards are discon-
nected.
2) Check if the shading correction plate is dirty.
3) Check if the conductor pattern on the CCD board is short circuited or
open circuited.
4) Check if the conductor pattern on the SLG board is short circuited or
open circuited.
5) Replace the lens unit.
6) Replace the SLG board.
NO
1) Check if the connectors of the exposure lamp and inverter are disconnected.
2) Check the SLG board if the connector pin CN9 is disconnected and the harness is short
circuited or open circuited.
3) Check if the conductor pattern on the SLG board is short circuited or open circuited.
4) Replace the SLG board.
5) Replace the inverter.
6) Replace the exposure lamp.
Содержание e-STUDIO200L
Страница 6: ......
Страница 12: ......
Страница 36: ...e STUDIO200L 202L 230 232 280 282 SPECIFICATIONS ACCESSORIES OPTIONS SUPPLIES June 2004 TOSHIBA TEC 1 24 ...
Страница 296: ...e STUDIO200L 202L 230 232 280 282 ERROR CODE AND SELF DIAGNOSTIC MODE June 2004 TOSHIBA TEC 2 260 ...
Страница 400: ...e STUDIO200L 202L 230 232 280 282 ADJUSTMENT June 2004 TOSHIBA TEC 3 104 ...
Страница 554: ...e STUDIO200L 202L 230 232 280 282 TROUBLESHOOTING June 2004 TOSHIBA TEC 5 124 ...
Страница 678: ...e STUDIO200L 202L 230 232 280 282 REMOTE SERVICE June 2004 TOSHIBA TEC 8 24 ...
Страница 700: ...e STUDIO200L 202L 230 232 280 282 DATA CLONING with USB STORAGE DEVICE e STUDIO202L 232 282 June 2004 TOSHIBA TEC 9 22 ...
Страница 702: ...e STUDIO200L 202L 230 232 280 282 WIRE HARNESS CONNECTION DIAGRAMS June 2004 TOSHIBA TEC 10 2 ...
Страница 707: ......
Страница 708: ......