No. 2B771-004EN*M
2-5
WARNING:
16. Remove the ECG electrodes from the patient before using
devices such as electric scalpels, high-frequency therapy
equipment, electrostimulators, or electric defibrillators.
In addition, when using such devices, do not let
ultrasound transducers, PCG microphones, or pulse wave
sensors to come into contact with the patient. Doing so
may result in the patient receiving a burn injury or an
electric shock.
CAUTION:
1. To prevent electric shock, do not connect peripheral units
(such as a video printer or video recording unit) to an
external outlet. Peripheral units should be connected inside
the system. For the connection procedures, contact your
TOSHIBA service representative.
2. If any abnormality of the system is found as a result of
inspection, stop using the system and contact your
TOSHIBA service representative for repair.
3. Do not spill or spray liquids such as water onto the system
or peripheral units.
2.5 Chemical
Hazard
Observe the following instruction in order to protect patients and operators from
inflammation or poisoning by chemical substances.
WARNING:
Handling the cord on this product will expose you to lead, a
chemical known to the State of California to cause birth
defects or other reproductive harm.
Wash hands after handling.
2.6 Electromagnetic
Compatibility
(EMC)
Definition: Electromagnetic compatibility (EMC) refers to the ability to function without
causing electromagnetic interference (EMI) in other devices or systems, as
well as to a certain level of immunity to EMI from other devices or systems.
Observe the following precautions to ensure EMC.
WARNING:
The use of transducers and cables other than those
specified, with the exception of transducers and cables sold
by Toshiba Medical Systems Corporation as replacement
parts, may result in increased emissions or reduced system
performance.
Содержание Aplio 500 TUS-A500
Страница 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Страница 50: ...No 2B771 004EN M 7 3 2 Move the system to its installation position using the handle 3 Lock the casters ...
Страница 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Страница 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Страница 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Страница 231: ...No 2B771 004EN M A 1 ...