No. 2B771-004EN*M
9-10
NOTE:
1. A message recommending standard power OFF (shutdown) is displayed in
the following cases.
The continuous power ON time
*1
has exceeded 48 hours or the
cumulative power ON time
*2
has exceeded 3.5 days (84 hours).
Displayed message:
[Due to the length of elapsed time since Startup,
Shutdown is recommended.]
The system has been started up from Standby mode 7 times or more
continuously.
Displayed message:
[Overuse of resume. Shutdown is
recommended.]
2. In the following cases, a message is displayed and the Standby function
cannot be used. Turn the system power OFF in standard mode
(shutdown).
The continuous power ON time
*1
has exceeded 96 hours or the
cumulative power ON time
*2
has exceeded 5 days (120 hours).
Displayed message:
[Due to the length of elapsed time since Startup,
select Shutdown or Cancel.]
The system has been started up from Standby mode 10 times or more
continuously.
Displayed message:
[Overuse of resume. Select Shutdown or
Cancel.]
*1: This is the cumulative time for which the system has been used
continuously.
*2: When Standby mode is used, this is the cumulative time for which the
system power has been ON after the system power was last turned ON in
standard mode (the standby periods are not included).
3. The standby function cannot be used in the following cases. In such
cases, the message shown below is displayed. When the use of these
functions is completed or terminated, the Standby function becomes
effective again. Note that if any settings of the functions have been
changed, the system power should be turned OFF in standard mode
(shutdown).
When using a maintenance function or a preset function
Displayed message:
[You can only select shutdown from
maintenance menu.]
When a USB flash drive is connected
Displayed message:
[Standby is disabled while the USB flash drive is
connected.]
4. If the main power switch on the power panel is turned OFF or the system
power cable is disconnected while the system is in Standby mode, Standby
mode is canceled and the system is started up in standard power ON
mode. If the battery kit (option) is installed, Standby mode is maintained
even when the main power switch on the power panel is turned OFF during
Standby mode. If the battery level is low, however, the system is started
up in standard power ON mode when it recovers from the Standby status.
If the battery kit is installed, the battery is charged automatically while the
main power switch on the power panel is ON.
The battery is intended to maintain the memory that stores system
information required for the Standby function. The system cannot be
started up using the battery only.
Contact your TOSHIBA service representative for the battery kit.
Содержание Aplio 500 TUS-A500
Страница 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Страница 50: ...No 2B771 004EN M 7 3 2 Move the system to its installation position using the handle 3 Lock the casters ...
Страница 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Страница 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Страница 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...