No. 2B771-004EN*M
23-4
23.5 Accuracy of Measurement
Accuracy of Each Clinical Measurement Parameter
Parameter
Ranges
Accuracy
- Image Depth Scale
Up to 400 mm
<
5% or <1 mm, if below 20 mm
- Physio/Image Scales
Up to 10 s
<
3%
- Distance/Depth
Up to 400 mm
<
5% or < 1 mm, if below 20 mm
- Area
Up to 999 cm
2
<
10% or < 0.41 cm
2
, if below 4 cm
2
- Circumference
Up to 1800 mm
<
5% or < 1 mm, if below 20 mm
- Angle
0 to 180
<
1
- Volume
Up to 4200 cm
3
<
16% or < 1.3 cm
3
, if below 8 cm
3
- Velocity
Up to 2000 cm/s
<
5% of the full scale
- Time
Up to 10 s
<
3% of the full scale
- Heart Rate
20 to 300 bpm
<
3%
- Accel
Up to 20000 cm/s
2
<
8%
- M distance
Up to 280 mm
<
5% or < 1 mm, if below 20 mm
- Slope
Up to 1000 mm/s
<
8%
Range of Applicability of the Accuracy Specifications
These measurements will satisfy the accuracies stated above over the entire selected field of
view. The accuracy stated is either the measured value or the calculated value for worst case
conditions.
It should be noted, however, that the above values are not guaranteed for measurements
performed on stored images that use manual calibration or MPR images that have been created
based on data acquired using the 4D transducer. For details of manual calibration, refer to the
operation manual Measurements volume. For details of the 4D transducer, refer to the section
describing the 4D function in the operation manual Applications volume.
*
Содержание Aplio 500 TUS-A500
Страница 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Страница 50: ...No 2B771 004EN M 7 3 2 Move the system to its installation position using the handle 3 Lock the casters ...
Страница 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Страница 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Страница 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Страница 231: ...No 2B771 004EN M A 1 ...