No. 2B771-004EN*M
9-2
WARNING:
6. Do not connect the diagnostic ultrasound system to the
same power outlet as another device. Doing so may
cause the circuit breaker of the facility to trip, fuses to
blow, or a fire or electric shock to occur.
9.1 Connecting the Power Cable and the Protective
Grounding
(1) Connect the power plug to an outlet for medical equipment. By doing this, the
protective ground line is connected.
(2) Turn ON the main power switch on the power panel.
The
[STANDBY]
LED lights.
Systems with the main power switch on the rear