No. 2B771-004EN*M
19-7
NOTE:
1.
(
)
is displayed at the bottom of the monitor while the data is
being stored on the HDD. Panel operation is disabled during this period.
2. When the number of cardiac cycles is specified as the image storing period
but the R wave is not detected, the image is stored for the period (seconds)
corresponding to the specified number of cardiac cycles.
3. The image save time for Clips defines the hard limit beyond which image
saving is not possible. However, for certain types of images, the actual
maximum time may be less than this value.
4. When R-R is selected as the method for specifying the image storing
range, a heart rate of 50 bpm or higher is assumed. If the patient's heart
rate is less than 50 bpm, the image may not be stored for the specified
number of cardiac cycles.
5. Frame missing error in recording with the Clips function
If the system detects that image frames are missing, the error message
below is displayed and the images are not recorded. Release freeze
status and acquire the image again.
[Failed in storing data, some frames were missed. Please try again.]
6. When the volume of stored data with raw data exceeds approximately
300 MB during retrospective storage, one of the following messages is
displayed.
(a) When the HDD is set as the recording destination
Less than 500 MB:
[Large Raw Data. Storing will take 20 seconds
or
more.]
Less than 700 MB:
[Large Raw Data. Storing will take 30 seconds
or
more.]
Less than 1024 MB:
[Large Raw Data. Storing will take 40 seconds
or
more.]
1024 MB or more:
[Large Raw Data. Storing will take 50 seconds
or
more.]
(b) When a location other than the HDD is set as the recording destination
Less than 500 MB:
[Data will be stored to HDD only. Please send
to your destination later. Large Raw Data.
Storing will take 20 seconds or more.]
Less than 700 MB:
[Data will be stored to HDD only. Please send
to your destination later. Large Raw Data.
Storing will take 30 seconds or more.]
Less than 1024 MB:
[Data will be stored to HDD only. Please send
to your destination later. Large Raw Data.
Storing will take 40 seconds or more.]
1024 MB or more:
[Data will be stored to HDD only. Please send
to your destination later. Large Raw Data.
Storing will take 50 seconds or more.]
7. In prospective mode, the data is stored at the specified destination
regardless of the volume of data.
8. The Still Store operation and print operation are disabled during image
storage in prospective mode.
Содержание Aplio 500 TUS-A500
Страница 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Страница 50: ...No 2B771 004EN M 7 3 2 Move the system to its installation position using the handle 3 Lock the casters ...
Страница 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Страница 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Страница 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Страница 231: ...No 2B771 004EN M A 1 ...