No. 2B771-004EN*M
9-8
NOTE:
5. When
[Shutdown]
is selected in the
[Power Control]
dialog during
copying of patient data, transfer of patient data, or burning of patient data
onto a CD/DVD, the following dialog is displayed.
No.
Item
Function
<1>
[Continue]
The current process is terminated and the
shutdown process is started. If this is selected
during CD/DVD burning, the burn process is
canceled (the CD/DVD will be damaged) and the
shutdown process is started.
<2>
[After procs end]
The shutdown process is started after the current
process is completed.
<3>
[Save procs]
The current process is paused (resumed) and the
shutdown process is started. If this is selected
during CD/DVD burning, the burn process is
canceled (the CD/DVD will be damaged) and the
shutdown process is started.
<4>
[Cancel]
The dialog is closed and the current process is
continued.