No. 2B771-004EN*M
4-1
4. Use Conditions
4.1 Power and Environmental Requirements
Item
Specifications
Power Line
voltage USA 120
VAC
10%
Europe 220 to 240 VAC
10%
Other 1 110 to 120 VAC
10%
Other 2 220 to 240 VAC
10%
Line frequency
50 Hz to 60 Hz
Power consumption
USA
1440 VA
Europe 1500
VA
Other 1 1440 VA
Other 2 1500 VA
Operating
environmental
conditions
Ambient temperature
10°C to 35°C
Relative humidity
35% to 80% (no condensation)
Atmospheric
pressure
700 hPa to 1060 hPa
Storage and
transportation
conditions
Ambient temperature
-10°C to 50°C
Relative humidity
30% to 80% (no condensation)
50% or less if the ambient temperature exceeds
40°C
Atmospheric
pressure
700 hPa to 1060 hPa
Patient environment
This system is designed to be used in the
environment specified in the figure below.