- 40 -
2
Input a desirable value according to “How to input a numerical value in the function mode”.
The value which can be inputted is 50 -120000 ms
Setting method
☞
’3.1.3 Setting values’
*FIX
INTEG
Setting*
Integ[ms]=
100
Shutter
=
*******
3
Press the [ENTER] switch to accept the integral time value on the display.
4
Pres the [ROTATE] switch to select inner-filter type. Each time the [ROTATION] switch is pressed,
the inner-filter type is changed as follows. Press the [ENTER]switch to accept the display value.
*******
↓
FILTER1
↓
FILTER2
↓
FILTER3
↓
FILTER4
↓
FILTER5
*******:
Inner-filter and integral time setting is reset. When conduct measurement after
selecting this, former inner-filter and integral time setting are erased and new
optimum setting are calculated.
Remember!
・
If inner-filter and integral time are not proper as follows, the accuracy in
measured data may be affected.
・
Designated integral time is smaller than proper integral time
・
Designated inner-filter is not proper
Remember!
・
A combination of following measuring angle and inner-filter can not be
used.
Angle
Inner-filter
0.2°
FILTER5
0.1°
FILTER4
、
FILTER5