4.1.1 RM and LM Command
By these commands, the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR is set to remote status (remote
mode) or stand alone status (local mode).
RM: Sets the remote mode.
LM: Sets the local mode.
When the power is turned on, the local mode is set. When the instrument operates for
communication, send the "RM" command first and then change to the remote mode.
PC
Command communication direction
Measuring instrument
“RM”+(Cr•Lf)
“OK”+(Cr•Lf)
“LM”+(Cr•Lf)
“OK”+(Cr•Lf)