- 101 -
5. Error Display
5.1 Error Display on the Instrument
Error display
Contents
Remedial measure
Attention!!
Temperature
Not
set
Please
Power
OFF
Because the instrument’s
inside is hot, the protection
circuit works.
Turn off the power.
Turn off the power first. Leave the
instrument in the environment
meeting the use condition (SR-3A:
0-30 °C, SR-LEDW/ SR-UL2/SR-
UL1R:0-35 °C) for about 30 minutes
and then turn on its power.
Attention!!
Low
Battery
Please
Power
OFF
The power of the backup
battery is low. Turn off the
power.
The power of the measurement
data backup battery is low and so
the data stored in the instrument
are deleted. It is necessary to
replace the battery. Consult
TOPCON TECHNOHOUSE or your
retailer.
***
ERROR
***
E001
Over
Range
hit
any
key
The measuring range is
exceeded.
Hit any key.
Lower the brightness of the object,
set a smaller measuring field or
insert the neutral density filter.
***
ERROR
***
E004
Sync
Error
hit
any
key
Can not receive the external
sync. signal.
Hit any key.
The status of receiving the external
sync. signal is abnormal. It is
necessary to check the sync.
signal generator or the connection
of BNC cable.
***
ERROR
***
EXXX
Comm
E
rror
E005 ~ E007
The communication error
occurred.
There is an communication error.
Check the communication
condition.
***
ERROR
***
E9XX
XXXXXXXXXXXXX
hit
any
key
E900 ~ E999
The internal error occurred.
Hit any key.
There is an internal error. Contact
TOPCON TEHCNOHOUSE or your
retailer.
•
If an error is still displayed though you take the above remedial measure, it is necessary to repair the
instrument. Contact TOPCON TECHNOHOUSE or your retailer.