![Topcon SR-3AR Скачать руководство пользователя страница 95](http://html1.mh-extra.com/html/topcon/sr-3ar/sr-3ar_instruction-manual_1128666095.webp)
- 84 -
4.1.20 FLD1, FLD2, FLD3 and FLD4 Command
Changes the measuring field setting in the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR.
FLD1: Change the measuring field to the 2.0 position.
FLD2: Change the measuring field to the 1.0 position.
FLD3: Change the measuring field to the 0.2 position.
FLD4: Change the measuring field to the 0.1 position.
PC
Command communication direction
Measuring instrument
FLD1/FLD2/FLD3/FLD4+(Cr
・
Lf)
OK+(Cr
・
Lf)
4.1.21 FLDR Command
Reads the current measuring field setting from the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR.
PC
Command communication direction
Measuring instrument
FLDR+(Cr
・
Lf)
OK+(Cr
・
Lf)
(Cr
・
Lf)
END+(Cr
・
Lf)
4.1.22 NL Command
Sets the integral time for the integral time delay function in the SR-LEDW/SR-UL2/SR-UL1R/SR-
3AR.
NL_####
:
The sign (#) is the integral time from 50 to 3 000 ms.
PC
Command communication direction
Measuring instrument
NL+(Cr
・
Lf)
OK+(Cr
・
Lf)
Содержание SR-3AR
Страница 1: ...INSTRUCTION MANUAL SPECTRORADIOMETER Rev 20...
Страница 2: ......
Страница 104: ...93...