- 106 -
Communication time
*Measurement is conducted under following condition
・
Object
White LED
・
Luminance
About3,000cd/m2
・
Measurement angle
2°
・
Integral time
100ms
・
Measurement mode
FIX
・
Measurement speed mode
HIGH SPEED
・
RS-232C parameter
9600bps_7bit_ODD_
1
bit
Temperature characteristic
Within ±3 % against luminance value
(The luminance of 20 °C should be standard in the range of 5°C ~30 °C.)
Polarized light characteristic
Luminance 1% or less Spectral radiance 2% or less(400~780nm)
Warm-up time
40 minutes or above
No warm-up time needed under following conditions.
Measuring field:2
Luminance of object to be measured is 1cd/m2 or above
Operating conditions is 23
°C
Calibration standard
TOPCON TECHNOHOUSE calibration standard (Standard light A / 23° C ± 3°C / 50 % ± 15 %RH)
Display
Dot matrix LC display (20 characters × 4 lines) with lighting
Interface
USB 2.0 (Full speed: 12 Mbps)
RS-232C
Communication speed:
4800/9600/19200/38400 BPS
Data length:
7 bit/8 bit
Parity:
EVEN/ODD/NONE
Stop bit:
1 bit/2 bit
Power supply:
Provided AC adapter
Remote via USB device
STB
Command
Spectral radiance mode
Approx. 1 sec
Remote via USB device
ST Command
Spectral radiance mode
Approx. 2.8 sec
Remote via USB device
Colorimetry mode
Approx. 0.6 sec
Remote via RS232C device
Colorimetry mode
Approx. 0.8 sec
Remote via RS232C device
Spectral radiance mode
Approx. 10.7 sec
*Communication time is depend on measurement
condition