- 91 -
4.1.36 CMF Command
Sets the CIE Color Matching Function (Type) setting in the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR.
CMF_#: The pound sign (#) is the CIE Color Matching function setting, either
CIE1931 or CIE170-2:2015.
For the pound sign (#), use 0 for a CIE1931 and 1 for a CIE170-2:2015.
PC
Command communication direction
Measuring instrument
CMF_0 or 1+(Cr
・
Lf)
OK+(Cr
・
Lf)
4.1.37 CMFR Command
Reads the CIE Color Matching function (Type) setting from the SR-LEDW/SR-UL2/SR-UL1R/
SR-3AR.
PC
Command communication direction
Measuring instrument
“CMFR”+(Cr
・
Lf)
“OK+(Cr
・
Lf)
CIE Color Matching function (Type)+(Cr
・
Lf)
“END” + (Cr
・
Lf)