![Topcon SR-3AR Скачать руководство пользователя страница 126](http://html1.mh-extra.com/html/topcon/sr-3ar/sr-3ar_instruction-manual_1128666126.webp)
- 115 -
* The indicated accuracy is the value when shipped.
Measuring
field
Measuring speed
High speed 2 mode
High speed 3 mode
Accuracy ensuring range
(against the standard light A)
2°
0.01~ 3 000 cd/
㎡
0.1~ 3 000 cd/
㎡
1°
0.03~ 9 000 cd/
㎡
0.3~ 9 000 cd/
㎡
0.2°
0.75~ 70 000 cd/
㎡
7.5~ 70 000 cd/
㎡
0.1°
3 ~ 300 000 cd/
㎡
30 ~ 300 000 cd/
㎡
Accuracy
Wavelength
-
±
0.3 nm (against the specified bright line of mercury)
Luminance
*1
Every
measuring
angle
±2%
Chromaticity
*1
2°
xy ±0.002
xy
±
0.003(0.1~0.5 cd/
㎡
)
xy
±
0.002(0.5~ cd/
㎡
)
1°
xy
±
0.003(0.3~1.5 cd/
㎡
)
xy
±
0.002(1.5~ cd/
㎡
)
0.2°
xy
±
0.003(7.5~40 cd/
㎡
)
xy
±
0.002(40~ cd/
㎡
)
0.1°
xy
±
0.003(30 ~150 cd/
㎡
)
xy
±
0.002(150~ cd/
㎡
)
Repeatability
Luminance
*2
2°
1.5%(0.01~0.05 cd/
㎡
)
0.4%(0.05~1 cd/
㎡
)
0.3%(1~ cd/
㎡
)
1.5%(0.1~0.5 cd/
㎡
)
0.4%(0.5~10 cd/
㎡
)
0.3%(10~ cd/
㎡
)
1°
1.5%(0.03~0.15 cd/
㎡
)
0.4%(0.15~3 cd/
㎡
)
0.3%(3~ cd/
㎡
)
1.5%(0.3~1.5 cd/
㎡
)
0.4%(1.5~30 cd/
㎡
)
0.3%(30~ cd/
㎡
)
0.2°
1.5%(0.75~4 cd/
㎡
)
0.4%(4~75 cd/
㎡
)
0.3%(75~ cd/
㎡
)
1.5%(7.5~ 40 cd/
㎡
)
0.4%(40~750 cd/
㎡
)
0.3%(750~ cd/
㎡
)
0.1°
1.5%(3~15 cd/
㎡
)
0.4%(15~300 cd/
㎡
)
0.3%(300~ cd/
㎡
)
1.5%(30~150 cd/
㎡
)
0.4%(150~3000 cd/
㎡
)
0.3%(3000~ cd/
㎡
)
Chromaticity
*3
2°
0.005 (0.01 ~0.05 cd/
㎡
)
0.0015(0.05~1 cd/
㎡
)
0.0005(1~ cd/
㎡
)
0.005 (0.1~0.5 cd/
㎡
)
0.0015(0.5~10 cd/
㎡
)
0.0005(10~ cd/
㎡
)
1°
0.005 (0.03~0.15 cd/
㎡
)
0.0015(0.15~3 cd/
㎡
)
0.0005(3~ cd/
㎡
)
0.005 (0.3~1.5 cd/
㎡
)
0.0015(1.5~30 cd/
㎡
)
0.0005(30~ cd/
㎡
)
0.2°
0.005 (0.75~4 cd/
㎡
)
0.0015(4~75 cd/
㎡
)
0.0005(75~ cd/
㎡
)
0.005 (7.5~40 cd/
㎡
)
0.0015(40~750 cd/
㎡
)
0.0005(750~ cd/
㎡
)
0.1°
0.005 (3~15 cd/
㎡
)
0.0015(15~300 cd/
㎡
)
0.0005(300~ cd/
㎡
)
0.005 (30~150 cd/
㎡
)
0.0015(150~3000 cd/
㎡
)
0.0005(3000~ cd/
㎡
)
*1 Against the standard light A and color matching function type CIE1931.
*2 For continuous 10 measurements, standard deviation is calculated twice.
*3 For continuous 10 measurements, max. value - min. value.