![Topcon SR-3AR Скачать руководство пользователя страница 108](http://html1.mh-extra.com/html/topcon/sr-3ar/sr-3ar_instruction-manual_1128666108.webp)
- 97 -
Initial bite
Function
Data type
Size
Qty
of
element
Remarks
2450
Wave length
Unsigned
integer
2
1
780 (fix)
2452
Spectral Radiance
Floating point
4
1
Spectral
Radiance
at
780nm
2456
“END”+Cr+Lf Fix
Character
code
1
5
Data size
2460B
3
Data (When error happen
)
Initianl bite
Function
Data type
Size
Qty
of
element
Remark
1
Error code
Character
code
1
4
5
“END”+( Cr
・
Lf)
Character
code
1
5
Data size
9
Error code
☞
’5.2 Error Code in communication’
(4)
“STCT” Command
When the measurement is conducted by order of STCT command, the measured data is returned with
comma separated text data from the instrument as follows;
Type
Output data
Remarks
1 1.490E+02,0.4458,0.4073+(Cr
・
Lf)
Luminance, Chromaticity x, y
“END”+(Cr
・
Lf)
Data end command
2 1.490E+02,0.2549,0.5240+(Cr
・
Lf)
Luminance, Chromaticity u’, v’
“END”+(Cr
・
Lf)
Data end command
3 1.631E+02,1.490E+02,5.374E+01+(Cr
・
Lf) Tristimulus values X, Y, Z
“END”+(Cr
・
Lf)
Data end command
4 1.631E+02,1.490E+02,5.374E+01,
Tristimulus values X, Y, Z,
0.4458,0.4073+(Cr
・
Lf)
Chromaticity x, y
“END”+(Cr
・
Lf)
Data end command
5 1.631E+02,1.490E+02,5.374E+01,
Tristimulus values X, Y, Z,
0.2549,0.5240+(Cr
・
Lf)
Ch
romaticity u’, v’
“END”+(Cr
・
Lf)
Data end command
6 2882,0.0002+(Cr
・
Lf)
Correlated color temperature (K), Deviation
“END”+(Cr
・
Lf)
Data end command
7 2.141231E-04,2.420037E-04
・・・・・
,
Spectral radiance
4.325765E-03,4.294558E-03+(Cr
・
Lf)
(380nm to 780nm)
“END”+(Cr
・
Lf)
Data end command
8 554,
Wavelength position,
4.294558E-03+(Cr
・
Lf)
Maximum Spectral radiance
“END”+(Cr
・
Lf)
Data end command
*When measurement is done in the colorimetry mode (D1 STCT), the data end command "END"
returned.