TH2683A Operation Manual
Ver.1.3
50
The :PWIDth command is usd to set the Handler bin output pulse width of the instrument.
The :PWIDth? query returns to the current Handler bin output pulse width.
Command syntax: COMParator: PWIDth <value>
Where,
<value> is any integer value among the range (1
‐
25ms) of Handler bin output pulse width.
For example: WrtCmd (“FUNCtion:PWID 10”); set the Handler bin output pulse width as 10ms..
Query syntax: FUNCtion: PWIDth?
Return format:<value><NL^END>
The command query returns to the current Handler bin output pulse width.
7.1.7
SYSTem Subsystem Commands
Command tree:
The :SYSTem:BEEP OFF (0) and ON (1) commands are used to set and query the system beeper of
the instrument.
For example:
SYST:BEEP
0
turn off the beeper
Query Syntax:
SYST:BEEP?
Return format: 0 or 1
SYSTem:VERSion?
It can query the current version number of the instrument, for example:
Ver 1.00
SYSTem:STSTus?
It can query the current status of the instrument.
Return format: DISCharging (indicating the syatem is in discharging), TESTing (indicating the
system is in test).
The SYSTem:HPOWer INTernal or EXTernalcommands are used to set the Handler interface
power.
Command syntax: SYSTem:HPOWer INTernal
set the HANDler power as INTernal.