TH2683A Operation Manual
Ver.1.3
38
7)
SCIP Command Reference
1. Data Conventions of This Manual
NR1
:
integer, for example:123
NR2
:
fixed number, for example: 12.3
NR3
:
floating number, for example: 12.3E+5
NL
:
CR character
,
integer: 10
^END
:
EOI (end) signal of IEEE
‐
488 bus.
7.1
TH2683A Subsystem Commands
●
DISPlay
●
TRIGger
●
COMParator
●
DISCharge
●
SYSTem
●
FUNCtion
●
FETCh?
●
Mass MEMory
NOTE: All the following commands relating to changing the instrument setup are
recommended to be used when the instrument is in discharge condition, otherwise it may result
in wrong results.
7.1.1
DISPlay Subsystem Commands
DISPlay subsystem commands are mainly used to set the instrument display page.
The :DISPlay? query returns to the current page.
Command Tree:
The :PAGE command sets the display page.
The DISPlay:PAGE? qurey returns to the current page.
Command Syntax: DISPlay:PAGE <page name>
Functions of <page name> are as follows
:
MEASurement
Set the display page to measurement display.
MSETup
Set the display page to measurement setup.
LTABle
Set the display page to pickout setup.
SYSTem
Set the display page to system setup.