TH2683A Operation Manual
Ver.1.3
21
3.4.3
Range Selection
There are six range resistances in TH2683A. They constitute six range segments (refer to figure
3
‐
1 below).
Range
Input Internal Resistance
1(1mA)
1k
Ω
2(100uA)
10k
Ω
3(10uA)
100k
Ω
4(1uA)
1M
Ω
5(100nA)
10M
Ω
6(10nA)
100M
Ω
Figure 3
‐
1 Range resistances
When the RANGE SET of the instrument is set as AUTO, for different resistances being measured,