A
Using the Direct Sample Probe
Running the Sequence
138
TSQ Quantum XLS Series User Guide
Thermo Scientific
Figure 68.
Probe at the first stop on guide bar
3. Click
OK
. The forepump evacuates the inlet valve.
4. When the Safe to Open the Valve message appears (
), click
OK
.
Figure 69.
Safe to open the valve message
5. Pull the inlet valve lever up to open the inlet valve.
6. Slide the probe into the vacuum manifold until the tip of the probe is fully inserted into
the ion volume holder. See
.
Inlet valve
knob
First stop on guide bar
Probe