6
Maintenance
Cleaning Ion Source Components
90
TSQ Quantum XLS Series User Guide
Thermo Scientific
Removing the L1, L2, L3 Lens Assembly and the Heated Q0 Quadrupole Assembly
The ion source includes the L1, L2, L3 lens assembly and the heated Q0 quadrupole
assembly. You must remove and disassemble the L1, L2, L3 lens assembly to clean the lenses.
A retainer clip fastens the L1, L2, L3 lens assembly and the heated Q0 quadrupole assembly to
the ion source block.
To remove the ion source lens assemblies
1. Wearing clean, lint- and powder- free gloves, unscrew the threaded hexagonal standoff
and remove the retainer clip that secures the lens assembly to the ion source block. See
2. Remove the heated Q0 quadrupole assembly and the L1, L2, L3 lens assembly from the
heater cartridges of the ion source. Place them on a clean surface.
Figure 52.
L1, L2, L3 lens assembly and heated Q0 quadrupole assembly removed from the ion source
L1, L2, L3 lens
assembly
Heated Q0 quadrupole
assembly
Retainer clip
Ion source
Threaded standoff