6
Maintenance
Completely Disassembling and Reassembling the Ion Source
98
TSQ Quantum XLS Series User Guide
Thermo Scientific
5. Reinstall the ion source into the vacuum manifold and reinsert the capillary column into
the ion source. See
.
6. Restart the system. See
“Starting Up the System after a Complete Shutdown”
Cleaning Stainless Steel Parts
IMPORTANT
For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor and the heater block.
Tools needed
• Acetone, reagent grade (or other suitable polar solvent)
• Aluminum oxide abrasive, number 600 (P/N 32000-60340)
• Applicators, cotton-tipped (P/N 00301-01-00015)
• Beaker, 450 mL
• Clean, dry gas
• Deionized water
• Detergent (Alconox™, Micro, or equivalent)
• Dremel™ rotary tool or equivalent (recommended)
• Foil, aluminum
• Forceps (P/N 76360-0400)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Glycerol, reagent grade
• Lint-free cloth
• Protective eyewear
• Tap water
• Toothbrush, soft
• Ultrasonic cleaner
Frequency
As needed to clean these stainless steel parts:
• Ion volumes
• Ion source block
• Ion source lenses
WARNING
Material and Eye Hazard
.
Wear impermeable laboratory gloves and eye
protection when performing cleaning procedures.
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......