7
Diagnostics and Troubleshooting
Troubleshooting
118
TSQ Quantum XLS Series User Guide
Thermo Scientific
Issue:
Excessive
injection port septum
bleed (Typical
m/z
207,
429, 355, 281).
Possible causes/solutions
• The septum is worn out or damaged. Replace the septum.
• Small pieces of septum are in the injection port liner. Replace the injection port liner;
condition the capillary column.
Issue:
Phthalate
background (
m/z
149,
167, 279).
Possible causes/solutions
• Phthalate contamination has occurred due to sample handling or solvent contamination.
• Packaging materials could be the source of phthalates.
• Isolate source of phthalates such as vial lids or plastic solvent containers and remedy.
Issue:
Excessive
hydrocarbon
contamination (Typical
ions are
m/z
43, 57, 71,
85, 99).
Possible causes/solutions
• Carrier gas tubing is contaminated.
• Isolate source of hydrocarbon contamination and remedy.
• Replace carrier gas tubing.
• Change carrier gas filters.
Issue:
Chemical
background due to
rhenium oxide
(
m/z
185/187, 201/203,
217/219, 233/235,
250/252).
Possible causes/solutions
• These series of rhenium oxide ions come from oxidation of the rhenium filament wire
due to introducing air into the ion source while the filament is on.
• Check for air leaks and remedy. For more information, see
Issue:
Spectra are
observed due to the
following solvents:
Solvent spectra
Acetone (
m/z
43, 58, 59)
Hexane (
m/z
41, 43, 56, 57, 58, 85, 86)
Methanol (
m/z
31)
Methylene chloride (
m/z
84, 83)
Toluene (
m/z
91, 92)
Trichloroethane (
m/z
151, 153)
Xylene (
m/z
105, 106)
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......