Tuning and Calibrating
Displaying the FC-43 Mass Spectrum
Thermo Scientific
TSQ Quantum XLS Series User Guide
63
Figure 33.
Define Scan dialog box
5. Click
Apply
to apply the settings.
6. To turn on the flow of calibration gas into the ion source, click the
Calibration Gas
button in the EZ Tune toolbar.
7. To turn on the filament, click the
Filament
button in the EZ Tune toolbar to turn on the
filament.
8. Choose
Control > On
to start the mass spectrometer scanning.
9. Observe the mass spectrum. If your mass spectrum looks very different from the one in
“Diagnostics and Troubleshooting”
Cal Gas Off
Cal Gas On
Filament Off
Filament On