Using the Direct Sample Probe
Removing the Probe
142
TSQ Quantum XLS Series User Guide
Thermo Scientific
Figure 74.
Raw file displayed in Qual Browser
Removing the Probe
Y
To remove the probe
1. Withdraw the probe until the guide ball on the probe reaches the first stop on the guide
2. Close the inlet valve by pushing the lever down.
3. Loosen the inlet valve knob by turning it counterclockwise.
CAUTION
Do not withdraw the probe beyond the point where the guide ball reaches
the first stop in the guide bar. Close the inlet valve first. Otherwise, the system vents
to the atmosphere.