D
Setting Up Sample Introduction Techniques
Setting Up H-ESI or Low Flow H-ESI Direct Infusion
134
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
1. With sample in the syringe, follow the procedures in
Syringe with the Syringe Pump.”
). The syringe pump is not shown.
2. Insert the grounding union (2) into the grounding union holder (3).
3. Use PEEK fittings to connect the red PEEK tubing (5) to the grounding union (2) on the
H-ESI spray insert (4).
4. Use a PEEK fingertight fitting to connect another length of red PEEK tubing (6) with the
syringe union adapter (1).
5. Use a PEEK fighertight (not shown) fitting to connect the other end of the red PEEK
tubing (6) to the grounding union (3).
Figure 69.
H-ESI direct infusion
IMPORTANT
To prevent the EMRS from degradation, observe the following precautions:
• Do not return any EMRS back to the original vial. Once the EMRS is out of the vial,
either use it or discard it.
• Use acetonitrile as a rinse solvent.
• Do not store the EMRS in a glass syringe. You may transfer the EMRS in a glass
syringe, but always discard any unused EMRS and rinse the syringe with acetonitrile.
• If you observe degradation products in the mass spectrum of the EMRS, rinse
everything in the sample line with acetonitrile (the fittings, syringe, tubing, and
union).
• Take these storage precautions for calibration and reserpine solutions: Refrigerate the
containers after opening. For long-term storage, keep refrigerated at 2–8 °C.
1
2
3
6
4
5
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...