Principles of Operation
Overview of the MS System
6
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
No.
Description
Function
Inlet Region
1.
Inlet: autosampler (optional), LC pump (optional),
Syringe pump, divert/inject valve
The sample enters the MS from the inlet.
Mass Spectrometer
2.
API source: H-ESI, APCI, or APPI
Generates ions.
3.
Vacuum system
Creates a vacuum for regions 4, 5, 6, and 7.
4.
Ion optics
In the ion optics and mass analyzer, the scan
functions manipulate the ions.
5.
Mass analyzer
6.
Ion detection systems
As the ions strike the dynode detector, they release a
cascade of electrons. These electrons generate a
voltage signal.
7.
Instrument control electronic assemblies
Control the MS components.
Data System
8.
Monitor
The data system processes the voltage signal and
displays it as a graph that can be printed.
9.
Computer
10.
Printer (optional)