14
Connecting the Sample Flow to the Source
Connecting the Sample Flow to the APCI Source
52
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Figure 23.
Connection to the H-ESI or the low-flow H-ESI source
1. Use red PEEK tubing (1) (with a fingertight fitting) to connect the sample flow to the
grounding union (2
).
2. Use red PEEK tubing (3) (with a fingertight fitting) to connect to the other end of the
grounding union (2).
3. Use a fingertight fitting to connect the other end of the red PEEK tubing (3) to the
H-ESI or the low-flow H-ESI spray insert (4).
Connecting the Sample Flow to the APCI Source
Connect the APCI spray insert to the sample introduction method. If you have the inlet
plumbed for H-ESI, you can leave the grounding union in place. You will not use it.
• Use tubing (red PEEK or natural PEEK (nano-Viper)) with a fingertight fitting to
connect the sample flow to the sample inlet of the APCI spray insert (
).
No.
Description
1.
Red PEEK tubing from the sample flow
2.
Grounding union
3.
Red PEEK tubing to the H-ESI or the low-flow H-ESI spray insert
4.
H-ESI or low-flow H-ESI spray insert
3
4
1
2
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...