23
Maintaining the API Source Interface
Removing the Ion Sweep Cone and the Ion Transfer Tube
104
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Removing the Ion Sweep Cone and the Ion Transfer Tube
Because buffer salts or high concentrations of samples can cause blockages in the bore of the
ion transfer tube, you must clean it.
If pressure in the ion transfer tube and RF lens region (Source Pressure gauge) drops below
2 Torr (TSQ Altis) or 1 Torr (TSQ Quantis or TSQ Fortis), a blocked ion transfer tube is
probably the cause.
1. Turn off the liquid flow to the API source.
2. In the Tune application, place the MS in
Standby
mode.
3. In the Ion Source pane, set the Ion Transfer Tube Temperature and Vaporizer
Temperature to
50 °C
or less and observe the readback temperatures.
4. Turn the system off.
Figure 53.
Ion sweep cone removed from the MS mount assembly
CAUTION Hot surface.
The external surface of the spray insert, API source housing,
and entry to the ion transfer tube can be hot enough to burn your skin. Allow the
parts to cool to room temperature (approximately 20 minutes) before you touch it.
No.
Description
No.
Description
1.
Release lever for the API source
interface
2.
API cone seal
3.
Ion sweep cone screw
4.
Ion sweep cone
1
2
4
3
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...