Using the MS Power Options
Shutting Down the MS System in an Emergency
32
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
The center of the selected power mode icon changes from white to green. The System
LED on the front panel turns yellow. To keep the API source clean, the MS reduces the
auxiliary and sheath gas flows to their standby default settings (2 arbitrary). The MS turns
off the electron multiplier, conversion dynodes, 8 kV power to the API source, main RF
voltage, and ion optic RF voltages. For a more complete list, refer to
Shutting Down the MS System in an Emergency
1. To turn off the MS in an emergency, turn off the main power switch (1 in
) on
the power panel on the right-side of the MS. (
This switch turns off all power to the MS, including the forepump, without harming
components within the instrument.
If you cannot safely reach the main power switch, unplug the power cord from the outlet.
2. To turn off the LC, autosampler, and data system computer in an emergency, use their
respective on/off switch or button.
Figure 19.
Main power switch on the power module
CAUTION
Emergency shut down procedure.
1
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...