Thermo Scientific
TSQ Altis, Quantis, and Fortis Hardware Manual
125
B
Setting Up the Syringe with the Syringe Pump
• Connecting the Syringe Union to the Syringe
Connecting the Syringe Union to the Syringe
Connect the syringe of the syringe pump to a syringe union.
1. Cut the ends of red PEEK tubing so that they are square (
). For best results, use
a polymeric tubing cutter. Poorly cut tubing can cause flow restrictions.
2. Make sure that the PEEK tubing is not crimped, kinked, or otherwise damaged.
3. Connect the PEEK tubing so that it contacts the bottom of the LC union’s 10-32, coned-
bottom receiving port. Tubing that is not properly seated can add dead volume to a
chromatographic system.
4. Tighten the PEEK fittings by only using your fingers. If you overtighten the PEEK
fittings, they can cause leaks.
Figure 64.
PEEK tubing connection to syringe union
No.
Description
No.
Description
1.
Red PEEK tubing
2.
Syringe Union (cross-sectional view)
3.
10-32 coned-bottom receiving port
4.
PEEK fingertight fitting
2
3
4
1