5
•
Installation
UHPLC System
Operating Manual
Page 77
Ensuring Proper Drainage of Waste
1. Position the waste container below the level of the system stack. To
prevent solvent vapors from evaporating back into the system, make
sure that the distance is sufficient.
2. Make sure the drain tubing is neither bent nor pinched.
3. To prevent liquid from flowing back into the system, the entire drain
tubing must remain below the drain port. Avoid the formation of
siphons.
TIP
To allow the waste liquid flowing off properly and avoid liquid
accumulation in the system base, the end of the waste line must not be
submerged in waste liquid (see figure below):
•
Shorten the tubing if required.
•
Observe the liquid level in the waste container. Empty the waste
container as necessary.
Figure 32: Free end of waste line in waste container
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...