8
•
Troubleshooting
Page 118
UHPLC System
Operating Manual
Pump problems
Possible Cause
Remedial Action
Piston seals in pump
leaking
Replace the seals (refer to
Pump Operating Manual
).
Air trapped in pump
head
Purge the pump (refer to
Pump Operating Manual
).
Baseline fluctuations
See "Periodic baseline fluctuations" in
.
Pump pulsation or flow
fluctuations
See "Pressure pulsation or inconstant pressure" in
Irreproducible gradient
Change the gradient.
Check the solvent line filters for permeability. Replace the filter frits if necessary
(refer to
Pump Operating Manual
).
Inappropriate detector settings
Possible Cause
Remedial Action
Improper wavelength,
e.g., in a UV spectrum
flank
Choose a detection wavelength that is located near the apex of the spectrum.
Detector response time
too short
Select a longer response time.
Detector response time
too long
Peaks are eventually not separated any more. Select a shorter response time.
Injection volume variation
Possible Cause
Remedial Action
Autosampler draws air
from vial
There is not enough amount of sample in the vial, the needle height setting is
incorrect, or there are too many replicates.
Air trapped in
autosampler flow path
Perform a wash cycle. Check that the injection valve is in Inject position and the
flow components of the autosampler are flushed with solvent.
Draw speed too high
Select a lower draw speed.
Gas content of sample
too high or saturated
Reduce the draw speed. Degas the sample if possible.
Injector needle clogged
or needle tip deformed
Replace the needle (refer to
Split Sampler Operating Manual
).
Piston seals in
autosampler metering
device leaking
Replace the metering device head (refer to
Split Sampler Operating Manual
).
Injection valve or other
autosampler
components leaking
Inspect the autosampler and the connections on the injection valve. Tighten leaking
connections. Replace leaking parts (refer to
Split Sampler Operating Manual
).
Needle seat worn out
Replace the needle seat (refer to
Split Sampler Operating Manual
).
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...