6
•
Operation
UHPLC System
Operating Manual
Page 85
6.2
Safety Guidelines for Operation
When operating the system, pay attention to the following safety
guidelines:
Observe all warning messages and precautionary statements presented
in
Observe all warning messages and precautionary statements presented
in the
Safety Guidelines for Operation
section in the
Operating Manual
for the individual modules of the Vanquish system.
NOTICE
Pay attention also to the following guidelines:
•
When operating the chromatography system, always set the lower
pressure limit for the pump. This prevents damage resulting from
leakage or from running the pump dry.
•
If there is evidence of leakage in the device, turn off the pump flow
and remedy the situation immediately.
•
If the pump flow is interrupted, act appropriately to protect the
components in the detector. For details, refer to the
Operating
Manual
for the detector.
•
Always verify that the autosampler is turned on before the pump
flow is on and pressure builds up. If the autosampler is turned off,
for example, after a power failure, stop the pump flow and wait until
the pressure is down to zero before turning on the autosampler or
other modules again.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...