8
•
Troubleshooting
Page 124
UHPLC System
Operating Manual
Post-column cooler does not reach set temperature
Possible Cause
Remedial Action
Post-column cooler not
installed correctly
Verify that the post-column cooler screws are tightened and the post-column
cooler is seated properly.
Autosampler does not reach set temperature
Possible Cause
Remedial Action
Ambient temperature
or humidity too high
Reduce the ambient temperature (for example, by ventilating the room), and/or
ambient humidity.
8.2.9
Power and Communication
Entire system has no power (all LED status bars unlit)
Possible Cause
Remedial Action
System power button
turned off
Turn on the system with the power button on the front left of the system base.
Single module has no power (LED status bar unlit)
Possible Cause
Remedial Action
Module power turned
off
Turn on the module with its main power switch.
Module not connected
to the mains
Connect the power cord.
Fuse has blown
Replace the fuses (refer to
Operating Manual
of the module).
Error in electronic
system
Contact Technical Support.
One or more modules cannot be turned off from the power button on the system base
Possible Cause
Remedial Action
System power
connection
interrupted
Check the system interlink connections between the system base and the system
modules, see
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...