8
•
Troubleshooting
Page 116
UHPLC System
Operating Manual
Possible Cause
Remedial Action
Capillaries clogged
Replace the capillaries.
Sample loop clogged
Replace the sample loop (refer to
Split Sampler Operating Manual
).
Solvent selector/
proportioning valve
defective
Contact Technical Support.
8.2.2
Ghost Peaks, Negative Peaks and Spikes
Ghost peaks
Possible Cause
Remedial Action
Contamination
(typically injector or
column)
Flush the system using an appropriate solvent.
Improper reference
wavelength
The sample must not absorb in the range of the reference wavelength. Consider
using a method without reference wavelength.
Selected bandwidth too
high
Select a lower bandwidth.
Degassing channels
contaminated
Purge the pump (all channels) to rinse the degassing channels (refer to
Pump
Operating Manual
).
Negative peaks
Possible Cause
Remedial Action
Improper reference
wavelength
The sample must not absorb in the range of the reference wavelength. Consider
using a method without reference wavelength.
Wrong polarization of
analog output
If you are using analog output, check the analog output polarization.
Spikes
Possible Cause
Remedial Action
Air trapped in flow cell
•
Check all fluid connections for tightness, in particular in the intake path.
•
Degas the mobile phase.
•
Install a restrictor at the flow cell outlet, observing the pressure specification of
the respective flow cell and the general guidelines for the flow cell.
Note
: Excessive backpressure will destroy flow cells!
Particles in flow cell
•
Flush the system (see
). If you suspect that a
capillary is contaminated with particles, in particular between the column and the
flow cell, replace the capillary.
•
Flush the flow cell with different flows if particles are trapped in flow cell.
•
Check if the column is defective.
UV/VIS lamp old,
defective, or not
properly installed
Check if the lamp is properly seated. Replace the lamp (refer to
Detector Operating
Manual
).
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...