8
•
Troubleshooting
UHPLC System
Operating Manual
Page 117
Possible Cause
Remedial Action
Electrical interferences
from other instruments
Avoid operation of equipment generating strong electric or magnetic fields in close
proximity. Do not connect equipment that may cause power disturbances to the
same power network used to supply the system.
Column temperature
significantly above
boiling point of mobile
phase
•
Install a restrictor at the flow cell outlet, observing the pressure specification of
the respective flow cell and the general guidelines for the flow cell.
Note
: Excessive backpressure will destroy flow cells!
•
Install a post-column cooler (refer to
Column Compartment Operating Manual
).
8.2.3
Peak Area Precision
Contamination or carry-over
Possible Cause
Remedial Action
Contamination in the
system
Flush the system using an appropriate solvent.
Autosampler carry-over
The needle may be contaminated. Wash the needle in the wash port. If carry-over
persists, replace the needle seat (refer to
Split Sampler Operating Manual
).
Environmental conditions
Possible Cause
Remedial Action
Unstable
environmental
conditions
Make sure that the temperature and air humidity are constant. Use column
thermostatting. Avoid draft. Use insulation on the capillary from the column
compartment to the detector.
Capillary problems
Possible Cause
Remedial Action
Capillary connections
not installed properly
or not tight, dead
volumes in capillary
connections
Properly install the capillary connections. Tighten the capillary connections. Replace
capillary.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...