8
•
Troubleshooting
Page 120
UHPLC System
Operating Manual
8.2.5
Pressure
Pressure pulsation or inconstant pressure
Possible Cause
Remedial Action
Air trapped in the
system
Purge the pump (refer to
Pump Operating Manual
).
Inlet and/or outlet
check valve
contaminated or
defective
Clean the inlet check valve or outlet check valve, and replace the check valve if
necessary (refer to
Pump Operating Manual
).
Pump inline filter (or
static mixer) clogged
Check the permeability of the inline filter (or static mixer). Replace the inline filter
(or static mixer) if necessary (refer to
Pump Operating Manual
).
Solvent line filter
clogged
Check the solvent line filters for permeability. Replace the filter frits if necessary
(refer to
Pump Operating Manual
).
Pump head
compression values not
reached
Check the compression values and take remedial action (refer to
Pump Operating
Manual
).
System clogged with
particles
Samples and insufficient solvent purity are frequently the source for particles
clogging the system. Follow the instructions for resolving clogging in the
autosampler (refer to
Split Sampler Operating Manual
).
High backpressure
Possible Cause
Remedial Action
One or more capillaries
in the system clogged
or damaged by bending
Check the capillaries in the system systematically from the detector to the pump.
Replace the capillaries as needed.
Pump inline filter (or
static mixer) clogged
Check the permeability of the inline filter (or static mixer). Replace the inline filter
(or static mixer) if necessary (refer to
Pump Operating Manual
).
Pre-heater clogged
Rinse the pre-heater capillary in reverse direction, using an appropriate solvent. If
necessary, replace the pre-heater (refer to
Column Compartment Operating
Manual
).
Column switching valve
clogged
Check the valve for indications of blockage. Replace the valve if necessary (refer to
Column Compartment Operating Manual
).
Post-column cooler
clogged or defective
Rinse the post-column cooler capillary in reverse direction using an appropriate
solvent. If necessary, replace the post-column cooler (refer to
Column
Compartment Operating Manual
).
Flow cell clogged
Clean the flow cell. Replace the flow cell if necessary (refer to
Detector Operating
Manual
).
Contamination in the
system
Flush the system with an appropriate solvent.
Low backpressure
Possible Cause
Remedial Action
Leak in system
Find and eliminate the leak.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...