8
•
Troubleshooting
UHPLC System
Operating Manual
Page 121
8.2.6
Retention Time Variation
Decreasing retention times
Possible Cause
Remedial Action
Improper mobile phase
composition
Check premixed mobile phase. Perform an OQ gradient accuracy test. A solvent
selector/proportioning valve may be defective. Contact Technical Support.
Increasing flow rate
Check flow rate settings. Perform OQ flow precision test.
Increasing retention times
Possible Cause
Remedial Action
Improper mobile phase
composition
Check premixed mobile phase. Perform an OQ proportioning test. A solvent
selector/proportioning valve may be defective. Contact Technical Support.
Decreasing flow rate
Check capillary connections for leaks. Check flow rate settings.
Piston seal leakage
Check the pump for piston seal leakage (refer to
Pump Operating Manual
).
Scattering retention times
Possible Cause
Remedial Action
Imprecise solvent
proportioning
•
Perform an OQ eluent proportioning test.
•
Clean the inlet check valve or outlet check valve, and replace the check valve if
necessary (refer to
Pump Operating Manual
).
•
A solvent selector/proportioning valve may be defective. Contact Technical
Support.
Pressure fluctuations
from pump
•
Purge the pump.
•
Check the compression values and take remedial action (refer to
Pump Operating
Manual
).
•
Check the pump for piston seal leakage (refer to
Pump Operating Manual
).
•
Check the solvent line filters for permeability. Replace the filter frits if necessary
(refer to
Pump Operating Manual
).
8.2.7
Baseline
High baseline drift
Possible Cause
Remedial Action
System not sufficiently
equilibrated
Flush the system until equilibration. Usually, a volume of 5–10 times the column
volume will be sufficient.
Unstable
environmental
conditions
1. Make sure that the temperature and the humidity are constant. You can
determine temperature fluctuations by recording the temperature channels.
2. Verify on the detector that the lamp house cover is properly installed and that the
front doors are closed. Avoid draft.
Flow cell contaminated
Clean the flow cell. Replace the flow cell if necessary (refer to
Detector Operating
Manual
).
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...