89
Thermo Fisher Scientific
Operation Manual HM550
Figure 93.
Inserting and replacing the filter unit.
[3]
[1]
[1]
[2]
• For this, remove the coarse filter as described above.
• Use lab gloves!!!
• Remove the coarse filter from the metal carrier by using forceps or
the like.
• Dispose of the coarse filter together with the section waste according
to the respective lab regulations.
Note:
e coarse filter is to be understood as a disposable filter and must not be
used again.
• Insert new coarse filter and close the filter.
Note:
In case section waste blocks the suction hoses of the Vacutome against ones
expectations, it can be removed by using the attached hose cleaner. During
this cleaning cycle, the Vacutome should be operated with maximum power
to guarantee that the released section waste is drained off properly.
Operating instructions
Option V – Vacutome