Cold D (Option DC)
To perform Cold D, the disinfectant tank (Fig. 112a.1) needs to be
attached to the back of the unit with two screws (Fig 112a). is only
needs to be done once.
• Position the tank on the bracket with the inlet hose centered in the
lower bracket.
• Insert the quick-action coupling onto the back of the unit, as shown
in Fig. 112a.2.
Once you have filled the tank you can utilize the Cold D process.
• Pour disinfectant solution (Sanosil®) into the disinfectant tank on
the back of the unit (Fig. 112a). Before filling the tank, check the
shelf life and the hazardous substance warnings on the disinfectant
(Sanosil®). Also, always follow the manufacturer’s instructions for
handling, storage and disposal of the disinfectant used (Sanosil®) as
well as the information provided by the manufacturer concerning
any precautionary measures to be taken in the event of an accidental
release or spillage of the disinfectant.
W
arning
Please note the safety warning in warnings on page 15.
• e max. capacity of the disinfectant tank is 1 l. is is typically
sufficient for approximately 5 Cold D cycles.
Preparing to perform
Cold D
Filling the tank with disinfectant
solution (Sanosil
®
)