• Turn the starter (fig. 135.5) in a counter-clockwise direction and
remove it from the socket.
• Insert the new starter into the socket.
• Turn the new starter clockwise until the starter is positively locked.
• Place the grip (fig. 135.1) onto the illumination ledge from above.
• Screw in the two red knurled screw clockwise from below and
tighten them.
• Plug in and then turn on the instrument and check the illumination.
Chapter 7
Conditions for the transportation
of the instrument
Repair or maintenance work are normally carried out at the site of
installation. If this is not possible for some special reasons, the instru-
ment can be returned to ermo Fisher Scientific Microm International. e
contact address can be found at the beginning of this operation manual.
• To guarantee trouble-free function of the instrument after transpor-
tation, please note the below-mentioned measures for the transpor-
tation preparation.
• In addition, the conditions for storage and transportation as mentioned
on pages 16 –18 must be observed during the entire transportation.
B
iohazard
!
Please also note the precautionary measures described in our safety pre-
cautions concerning biological hazardsꢞ
• Turn off the instrument.
• Unplug the unit.
• Remove blade and/or knife from blade and/or knife carrier.
• Remove section waste trays, brush shelf, blade/knife carrier and
other accessories as well as tools from the cryo chamber.
Taking back the instru-
ment for the repair or
routine maintenance
Measures for
closing down
Conditions for the transportation of the instrument
Taking back the instrument for repair or routine maintenance