Note:
e upper wooden cover now serves as an incline (fig. 5a.1) on which the
cryostat can be moved from the pallet to the floor by using the handle (fig. 5a.2)
• e hinges of the upper wooden cover (fig. 5.1) are inserted into
the screws of the bottom plate (fig. 5a.3). e correct side for this is
marked with arrows.
• e tool (fig. 5b.2, size 6) for the height adjustment of the setting
feet of the cryostat is packed together with the accessories of the
cryostat (separate carton on the foot rest of the cryostat).
• Turn the setting feet completely upwards via this tool so that the
cryostat stands with its rollers on the bottom plate (fig. 5a.3).
• Via the handle (fig. 5a.2) roll the cryostat from the bottom plate
(fig. 5a.3) over the wooden cover (fig. 5a.1). In doing so, the rear
side of the instrument must show to the incline (fig. 5a.1).
Figure 5b.
Allan Key SW 6 for hight adjustment of the cryostat.
Figure 5a.
Removing the bottom plate
[2]
[3]
[3]
[1]
[5]
Operating instructions
Setting up the cryostat