Operation Manual HM550
c
aution
!
e instrument is very heavy. While the instrument rolls down on the
incline, it must be guaranteed that the instrument does not move in an
uncontrolled way.
• Now the cryostat can be rolled to its site of installation.
• For possible return shipments, please keep the packing.
W
arning
e instrument must only be moved in standing or slightly tilted (ap-
prox. 30°) position.
Choose installation site that
• enough ventilation for the cooling system is guaranteed.
• the distance between wall and rear panel is approx. 15 cm. For this,
install the distance bolts (fig. 6a.1).
• the suction areas on either side are kept free (fig. 6a.2).
• the mains switch for separating the instrument from the power supply
is accessible any time.
Moreover, the installation site must be free from:
• draught by open doors or by air conditioning systems.
• direct exposure to sunlight into the cooling chamber.
[1]
[1]
Figure 6.
Balancing out the machine.
Operating instructions
Setting up the cryostat