8
|
Terms Definition
Section
Section
Thermo Scientific
Orion 2295 Phosphate Analyzer
|
57
Serial Number
Serial number of the instrument.
None
UNIT LABEL CHANGE
Label Name
Change the title of the instrument.
Maximum 9 characters A–Z, 0–9, / , - ,
and space are allowed
PRINTING OPTIONS
Print System
Configuration
Prints all the system parameters.
None
Print Analysis
Parameters
Prints only the analysis parameters.
None
Print Method
Parameters
Prints only the method parameters.
None
Print Clean
Parameters
Prints only the clean parameters.
None
Online Print
Prints the date, time and concentration after
each analysis cycle.
Enabled, Disabled
EDIT CALIBRATED VALUES
Pump1, Pump2
To change the flow rate of the pumps after
their calibration.
0.10 to 60.00 mL
Optics Dark
The level detected as dark light level.
0 to 300 mV
Optics Xref
The level used as reference potential for
light energy level.
0 to 2400.00 mV
PRIME FLUIDICS
Priming Volume
The volume of the fluid selected for priming.
0 to 99.9 mL
PUMPS & STIRRER
Pump1, Pump2
To check the functionality of pumps.
ON, OFF
Stirrer
To check the functionality of stirrer.
ON, OFF
RELAYS
Relay1, Relay2,
Relay3, Relay4
To check the functionality of relays.
ON, OFF
ANALOG OUTPUTS
Channel1,
Channel2
To test the functionality of analog outputs
for different selectable currents.
0 mA, 4 mA, 10 mA, 12 mA, 20 mA
LED & ADC
Select LED
To test the functionality of LED.
3, None
Intensity
Intensity of the LED.
0% to 100%
Optics Amplifier
gain
Changing the gain of the amplifier.
0 to 255
Terms
Description
Possible Values, Options