Orion 2295 Phosphate Analyzer
Thermo Scientific
Section
Section
Relays
Allows technician to selectively check the functionality of the four relays. The relay status is
shown next to the selected relay. Pressing the ENTER key toggles the state of the relay. The
relay initial status will be updated upon leaving this diagnostics menu.
Analog Output
Allows technician to selectively check the two analog outputs. Previous analog output setting will
be updated upon leaving this diagnostics menu.
LED and Detector Control
Allows the technician to check the functionality of the optics system and its individual
components. Also allows adjusting the detector preamp gain for the selected LED. Any new
setting of the gain or of the intensity of the LED will override any previous optics calibration
setting.
CONSULT FACTORY before overriding any setting.
This diagnostics menu allows direct reading of the mV level detected at the preamp and it shows
the corresponding absorbance calculated. The internal electronics enclosure temperature and
the reaction cell temperature are also displayed for reference.
Default and Lost or Forgotten Passwords
The default passwords are as follows:
Technician - 123***
User 1 -111111 (6 ones)
User 6 -666666 (6 sixes)
If a user password is forgotten, the technician can delete the user and then add the user back.
If the technician password is forgotten, call customer support: 1.800.225.1480.
Figure 11: UNLOCK keypad