9
|
Maintenance Schedule
62
|
Orion 2295 Phosphate Analyzer
Thermo Scientific
Section
Section
Figure 13: Sample chamber detail
Stand-By Mode and Extended Shutdown
To put the system in Stand-by mode, make sure to press the STOP key and check the DISPLAY
field next to F4 to show MANUAL RESTART. If AUTO RESTART is selected, you need to set the
mode to Manual:
MENU
➔
SETUP
➔
+ ANALYSIS PARAMETERS
➔
MODE
The system will switch to Stand-By after 60 minutes of no keypad activity. In Stand-By, the
analyzer will rotate the pumps periodically to avoid pinched tubing from sealing.
Extended Shutdown
To avoid clogging, flush the entire system with distilled water before shutdown. Disconnect and
empty out the reagents bottles and flush the lines using FLUIDICS PRIME menu. Release the
pressure on the peristaltic pump tubing.
Use the Table 3 to diagnose and correct analyzer problems.