2
|
Introduction
Section
Section
Thermo Scientific
Orion 2295 Phosphate Analyzer
|
7
Descriptions of the system hardware and user-interface software (the control panel) are
included, as an overview of the analytical techniques employed in the analyzer. Step-by-step
instructions guide you through all operating procedures and field programming.
The analyzer is designed for easy operation. Each analyzer in this family runs in a
fully automated mode for routine, day-to-day water analysis.
The need for operator interaction is minimal. Accordingly, the information required by the routine
operator of this analyzer is modest, and that information is provided in
Section 4 Startup and
Operation
, of this manual.
Important standard analyzer capabilities include:
Sampling the process stream automatically.
Calculating concentration of your sample and reporting it in units of concentration.
Recalibrating automatically at preprogrammed intervals to compensate for any changes in
captured sample volume, pump flow rates and reagent composition.
Cleaning automatically (when programmed) at user-specified intervals to prevent
precipitation build-up, coating of the optics or bacterial/fungal growth.
The analyzer brings colorimetric laboratory precision to the real world of unattended process
analyzing. It is designed for long life under continuous, rigorous use.
The Orion 2200 series analyzers are designed for accommodating several applications. Each
application requires a specific configuration in hardware and software that is set at the factory.
The product as supplied will have a configuration sheet detailing the setup and the detailed tests
and validation that were conducted at the factory. This analyzer model is intended to use for
clean water applications such as power plant
water analysis only, suspended solids must be
less than 60 microns.
User Interface
General Description
The user-interface consists of a keypad and a graphical LCD display. The keypad is located to
the right of the display. The "F1-F4" function keys (soft-keys) take on different meanings
depending on the menu selection. The column on the left of the soft-keys labels the keys with
their appropriate functions.
The RUN key starts the analysis and the STOP key stops the current functioning of the
instrument. The + and - keys allow for navigation between pages. The ENTER key confirms
selections made in the menus. Additional navigation is provided by the soft-keys (in appropriate
menus), for left / right or up / down cursor movement.