Installing Peristaltic Pump Head Tubing
Press the STOP key.
Disengage the U shape pressure plate of the pump head assembly by pressing with two fingers
and gently move upside as shown in
Figure 6
. Slide the tube fitting retainers at each side.
Replace the old tube by sliding off the barbed fitting. If necessary, clean the pump rollers.
Lubricate the tubing with a pea-sized quantity of silicone oil.
Fit the new tube and replace in the pressure plate. Note that the pressure plate sides must click
properly in the pump head assembly and align on both sides.
Repeat above steps for all pumps.
You can test the pump under Diagnostics menu for proper operation:
MENU
➔
SERVICES
➔
DIAGNOSTICS
➔
PUMPS & STIRRER
Figure 6: Installing peristaltic pump head
tubing - (Single channel)